Testing for Continuity
Testing for Continuity |
Special Tools
EL 39200 Digital Multimeter (DMM)For equivalent regional tools, refer to Special Tools .
Refer to Test Probe Caution
The following procedures verify good continuity in a circuit.
With a Digital Multimeter (DMM) |
1. |
Set the DMM to the Ω position.
|
2. |
Disconnect the power feed (i.e. fuse, control module) from the suspect circuit.
|
3. |
Disconnect the load.
|
4. |
Select the MIN MAX function on the DMM.
|
5. |
Connect one lead of the DMM to one end of the circuit to be tested.
|
6. |
Connect the other lead of the DMM to the other end of the circuit.
|
7. |
If the DMM displays low or no resistance, the circuit has good continuity.
|
With a Test Lamp |
Note | ||
Only use the test lamp procedure on low impedance power and ground circuits. |
1. |
Remove the power feed (i.e. fuse, control module) from the suspect circuit.
|
2. |
Disconnect the load.
|
3. |
Connect one lead of the test lamp to one end of the circuit to be tested.
|
4. |
Connect the other lead of the test lamp to battery positive voltage.
|
5. |
Connect the other end of the circuit to ground.
|
6. |
If the test lamp illuminates (full intensity), then the circuit has good continuity.
|