Vacuum pumping
Vacuum pumping |
Important | ||
Vacuum pumping of the system should be carried out on both the low-pressure and high-pressure sides simultaneously. A non-return valve in the inlet on the compressor's low-pressure side makes vacuum pumping on only one side impossible. |
||
1. |
Using a snap-on coupling, connect the blue low-pressure hose to the low-pressure side of the AC system.
|
2. |
Using a snap-on coupling, connect the red high-pressure hose to the high-pressure side of the AC system.
|
3. |
Open the valves on both snap-on couplings by screwing in the knobs.
|
4. |
Connect the yellow hose on the pressure-gauge stand's middle outlet to the refrigerant flask.
|
5. |
Open the cocks on the blue low-pressure hose and red high-pressure hose on the pressure-gauge stand.
|
6. |
Start the vacuum pump.
|
7. |
Open the vacuum valve and let the pump run for 30 minutes.
|
8. |
Close the vacuum valve when vacuum pumping is finished after about 30 minutes. The pressure should then be 0-10 mbar.
|
9. |
Switch off the pump.
|
10. |
Move the AC hoses slightly and make sure that there are no leaks in the system. The pressure must not increase by more than 10 mbar in five minutes.
|